| Share On: | |
|---|---|
| Categories: | New Products | Motion, Temperature and Leak Detection Sensors | CL & DL Series Environmental Data Logger |
| Review: | Write your review |
| Resources: |
The DL-301-IP65 is an IP65-rated industrial data logger for monitoring carbon monoxide (CO), temperature, relative humidity and calculated dew point. It uses an electrochemical CO sensor with a 0 to 1000 ppm measurement range and includes a 2.8-inch LCD touch screen. The module can store up to 450,000 time-stamped records for later download and analysis.
For system integration, the DL-301-IP65 provides one RS-485 port and a 10/100 Base-TX Ethernet port. It supports DCON and Modbus RTU over RS-485, plus Modbus TCP and MQTT over Ethernet. The module can be powered from a +12 to +48 VDC input or through IEEE 802.3af Class 1 PoE. Four PhotoMOS relay outputs can be used with audible or visual alarm devices or compatible control equipment.
The DL300 Utility can configure modules, monitor real-time values and organize multiple DL-300 devices into groups. Logged data can be downloaded and exported as a CSV file for analysis in spreadsheet or other compatible software.
Real-time data can be viewed using the DL300 Utility, the iOS or Android app, or a web browser when the monitoring device and data logger are connected to the same local network.
The display-message-on-screen function supports multiple language character sets based on UTF-8 encoding. Either pre-configured messages or dynamic messages can be remotely displayed using Modbus commands, or a dynamic message can be sent via the web-based interface.
| Software | |
|---|---|
| Built-in Web Server | Yes |
| CPU Module | |
|---|---|
| Real Time Clock | Yes |
| Watchdog Timer | Yes, Module (2.3 sec), Communication (Programmable) |
| Display | |
|---|---|
| Backlight Life | 20,000 hours |
| Brightness | 160 cd/m2 |
| Touch Panel | Yes |
| LCD Information Displayed | 2.8 TFT (Resolution 240 x 320 x 16), Defective Pixels <= 3 Backlight Life 20,000 hours |
| LED Indicators | |
|---|---|
| Status | PWR: Green for normal operation Link: Green for the Ethernet-linked S1 ~ S4: Red for an alarm condition |
| System | |
|---|---|
| Security | IP filter (whitelist) and Password (web) |
| Real Time Clock | Yes |
| Data Logger | Yes, up to 450,000 records |
| Temperature Measurement | |
|---|---|
| Range | -20 ~ +50°C |
| Resolution | 0.1°C |
| Accuracy | ±0.6°C |
| Humidity Measurement | |
|---|---|
| Range | 0 ~ 100% RH, Non-condensing |
| Resolution | 0.1% RH, Non-condensing |
| Accuracy | ±5% RH, Non-condensing |
| Dew Point | |
|---|---|
| Range | Calculated using temperature and relative humidity |
| Resolution | 0.1°C |
| CO Measurement | |
|---|---|
| Sensor Type | Electrochemical |
| Range | 0 ~ 1000 ppm |
| Resolution | 1 ppm |
| Accuracy | ±5% of measured value |
| Response Time | 30 seconds |
| Warm-up Time | 300 seconds |
| Relay Output | |
|---|---|
| Type | PhotoMos Relay 4, 100 VDC @ 1 A |
| Ethernet | |
|---|---|
| Ports | 10/100 Base-TX, 8-Pin RJ-45 x1(Auto-negotiating, Auto-MDI/MDIX, LED indicators) |
| PoE | Yes |
| Protocol | DCON, Modbus RTU, Modbus TCP, MQTT |
| COM Ports | |
|---|---|
| Ports | 1 x RS-485 |
| Baud Rate | 1200 ~ 115200 bps |
| Protocol | DCON, Modbus/RTU |
| Power | |
|---|---|
| Consumption | PoE: 2.9 W (Max.) Non-PoE: 2.7 W (Max.) |
| Powered from PoE | IEEE 802.3af, Class 1 (48 V) |
| Powered from Terminal Block | +12 ~+48 VDC |
| Mechanical | |
|---|---|
| Dimensions (mm) | 116 x 211 x 70 (W x L x H) |
| Installation | DIN-Rail or Wall Mounting |
| Ingress Protection Rating | IP65 |
| Environment | |
|---|---|
| Operating Temperature | -20 ~ +50°C |
| Storage Temperature | -30 ~ +75°C |
| Humidity | 10 ~ 90% RH, Non-condensing |